Technical Problems: Difference between revisions

From ECGpedia
Jump to navigation Jump to search
No edit summary
Line 1: Line 1:
{{authors|
|mainauthor= [[user:Drj|J.S.S.G. de Jong]]
|moderator= [[user:Drj|J.S.S.G. de Jong]]
|supervisor=
}}
==Lead reversals==
==Lead reversals==
[[Image:cableReversal1.png|thumb|Right and left arm lead reversal can be distinguished from the (much rarer) dextrocardia by looking at the p-wave.]]
[[Image:cableReversal1.png|thumb|Right and left arm lead reversal can be distinguished from the (much rarer) dextrocardia by looking at the p-wave.]]

Revision as of 22:20, 18 September 2007

Author(s) J.S.S.G. de Jong
Moderator J.S.S.G. de Jong
Supervisor
some notes about authorship

Lead reversals

Right and left arm lead reversal can be distinguished from the (much rarer) dextrocardia by looking at the p-wave.
Right arm and left leg lead reversal. Lead II now measures the signal between the left and right leg, which is remote from the heart.

Sometimes an ECG is made properly. Mistakes do happen and leads can be switched. Always remain careful to check this or you might come to the wrong conclusions. One of the most common mistakes is to switch the right and left arm. This will result in negative complexes in I, indicating a right axis deviation!

Common mistakes are reversal of:

  • right leg and right arm:
    • Hardly any signal in lead II.
  • right and left arm electrodes;
    • reversal of leads II and III
    • reversal of leads aVR and aVL
  • left arm and left leg:
    • reversal of leads I and II
    • reversal of leads aVR and aVF
    • inversion of lead III
  • right arm and left leg:
    • inversion of leads I, II and III
    • reversal of leads aVR and aVF


It is possible to distinguish lead reversal and dextrocardia by watching the precordial leads. Dextrocardia will show an R wave inversion, wheras lead reversal will not.

Artefacts

Movement artefacts
Increasing movement artefacts in a Parkinson patient. The patient was in sinus rhythm! (which doesn't show on this short recording)
Baseline drift. The amplifier in the ECG machine has to re-find the 'mean'. This often occurs right after lead connection and after electric cardioversion.
Cardioversion from atrial fibrillation to sinusrhythm, with clear baseline drift.
Electrical interference from a nearby electrical appliance. A typical example is a 100 Hz background distortion from fluorescent lights. Not to be confused with atrial fibrillation.
Another example of an artefact caused by an electrical appliance. The patients rhythm is regular. This strip shows 10 QRS complexes.

Artefacts (disturbances) can have many causes. Common causes are:

  • Movement
  • Electrical interference