* Farther to the right is the interpretation of the ECG written (this may be missing in a 'fresh' ECG, but later the interpretation of the cardiologist or computer will be added) | * Farther to the right is the interpretation of the ECG written (this may be missing in a 'fresh' ECG, but later the interpretation of the cardiologist or computer will be added) | ||
* down left is the 'paper speed' (25 mm/s on the horizontal axis), the sensitivity (10mm/mV) and the filter's frequency (40Hz, filters noise from eg. lights) | * down left is the 'paper speed' (25 mm/s on the horizontal axis), the sensitivity (10mm/mV) and the filter's frequency (40Hz, filters noise from eg. lights) | ||
* | * There is a calibration. At the beginning of every lead is a vertical block that shows with what amplitude a 1 mV signal is drawn. So the height and depth of these signals are a measurement for the voltage. If this is not the set at 10 mm, there is something wrong with the machine setting. | ||
* further we have the ECG leads themselves of course, these will be discussed below. | * further we have the ECG leads themselves of course, these will be discussed below. | ||
Note that the | Note that the layout is different for every machine, but most machines will show the information above somewhere. | ||