m (New page: R. Tukkie, R. Rienks, A.A.M. Wilde In 2002, an a trial demand inhibited (AAI) pacemaker was implanted in a young male (born 1984) because of a primary arrhythmia disorder. On a routine H...) |
mNo edit summary |
||
(2 intermediate revisions by the same user not shown) | |||
Line 1: | Line 1: | ||
R. Tukkie, R. Rienks, A.A.M. Wilde | {{NHJ| | ||
|mainauthor= '''R. Tukkie, R. Rienks, A.A.M. Wilde''' | |||
|edition= 2004:11,510 | |||
}} | |||
[[Image:Puzzle200_11_510.jpg|Figure 1|thumb]] | |||
In 2002, an a trial demand inhibited (AAI) pacemaker was implanted in a young male (born 1984) because of a primary arrhythmia disorder. On a routine Holter recording, the following rhythm strip was recorded early in the morning during sleep (figure 1). | |||
In 2002, an a trial demand inhibited (AAI) pacemaker was implanted in a young male (born 1984) because of a primary arrhythmia disorder. On a routine Holter recording, the following rhythm strip was recorded | |||
'''Which phenomenon is visible?''' | '''Which phenomenon is visible?''' | ||
Line 16: | Line 17: | ||
'''What is the consequence of the observed phenomenon in question 1 in the setting of this patient’s primary electrical disease?''' | '''What is the consequence of the observed phenomenon in question 1 in the setting of this patient’s primary electrical disease?''' | ||
As a result of the pause after the non-conducted atrial paced beat, the QT interval further prolongs significantly to QTc 740 ms, increasing the risk of torsades de pointes. QT prolongation at bradycardia is a feature of | As a result of the pause after the non-conducted atrial paced beat, the QT interval further prolongs significantly to QTc 740 ms, increasing the risk of torsades de pointes. QT prolongation at bradycardia is a feature of type III LQTS. | ||
type III LQTS. | |||
'''What action should be undertaken?''' | '''What action should be undertaken?''' | ||
It is mandatory to prevent relative bradycardia and pauses due to Wenckebach block in this patient. This can be achieved by adding a ventricular lead and programming the new pacemaker to duel-chamber with an AV delay long enough to advocate intrinsic AV conduction to the ventricle. Our patient underwent an uneventful upgrade and a control Holter recording showed excellent prevention of pauses. | It is mandatory to prevent relative bradycardia and pauses due to Wenckebach block in this patient. This can be achieved by adding a ventricular lead and programming the new pacemaker to duel-chamber with an AV delay long enough to advocate intrinsic AV conduction to the ventricle. Our patient underwent an uneventful upgrade and a control Holter recording showed excellent prevention of pauses. |