File:E0003147.jpg
Jump to navigation
Jump to search
Size of this preview: 800 × 306 pixels. Other resolution: 3,004 × 1,148 pixels.
Original file (3,004 × 1,148 pixels, file size: 473 KB, MIME type: image/jpeg)
Summary
Description |
This is a threshold test of a VVI pacemaker. Note the effect of the PVCs, the first of which competes with the pacer stimulus and suggests capture when there may be none. Also note that the second to last 0.6 volt pulse captures the ventricle but the last, with the same pulse width, does not. |
---|---|
Category | |
Source |
EKG World Encyclopedia http://cme.med.mcgill.ca/php/index.php , courtesy of Michael Rosengarten BEng, MD.McGill |
Date |
2012 |
Author |
Michael Rosengarten BEng, MD.McGill |
Permission |
Creative Commons Attribution Noncommercial Share-Alike License |
File history
Click on a date/time to view the file as it appeared at that time.
Date/Time | Thumbnail | Dimensions | User | Comment | |
---|---|---|---|---|---|
current | 22:48, 18 February 2012 | 3,004 × 1,148 (473 KB) | DarrelC (talk | contribs) |
You cannot overwrite this file.
File usage
The following page uses this file: