File:E0003108.jpg
Original file (3,004 × 673 pixels, file size: 683 KB, MIME type: image/jpeg)
Summary
Description |
The reason why the pacer was pacing at 75/min in the DDD mode was because there is a supraventricular tachycardia (note atrial activity, marked with the turquise arrows, moving through the ventricular complexes) at about 150/min and the DDD pacer was producing 2:1 A/V block. In tracing above with the pacemaker set to a VVI mode, the pacemaker no longer senses the atrium and paces at the set rate of 70/min. |
---|---|
Category | |
Source |
EKG World Encyclopedia http://cme.med.mcgill.ca/php/index.php , courtesy of Michael Rosengarten BEng, MD.McGill |
Date |
2012 |
Author |
Michael Rosengarten BEng, MD.McGill |
Permission |
Creative Commons Attribution Noncommercial Share-Alike License |
File history
Click on a date/time to view the file as it appeared at that time.
Date/Time | Thumbnail | Dimensions | User | Comment | |
---|---|---|---|---|---|
current | 21:56, 16 February 2012 | 3,004 × 673 (683 KB) | DarrelC (talk | contribs) |
You cannot overwrite this file.