File:E0003148.jpg
Original file (3,004 × 1,599 pixels, file size: 971 KB, MIME type: image/jpeg)
Summary
Description |
This is an EKG rhythm and marker strip from a patient with a Medtronic model 7108 Minuet dual chamber pacemaker. The patient had intact A/V conduction, and with a long paced A/V interval, conducted paced atrial beats to the ventricle. The atrial pacing threshold was determined by pacing at a rate faster than the patient's own rate and decreasing the atrial pulse width until atrial capture was lost, and ventricular pacing occurred (the upper tracing). In spite of returning the atrial pulse width back to the last successful value, ventricular pacing continued . Slowing the pacing rate terminated ventricular pacing (the lower tracing) and ventricular pacing did not return when the pacing rate was speeded up again. |
---|---|
Category | |
Source |
EKG World Encyclopedia http://cme.med.mcgill.ca/php/index.php , courtesy of Michael Rosengarten BEng, MD.McGill |
Date |
2012 |
Author |
Michael Rosengarten BEng, MD.McGill |
Permission |
Creative Commons Attribution Noncommercial Share-Alike License |