File:E339.jpg

From ECGpedia
Revision as of 10:44, 21 February 2012 by DarrelC (talk | contribs)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

Original file(3,004 × 599 pixels, file size: 646 KB, MIME type: image/jpeg)

Summary

Description

This is a tracing in taken from an 80 year old man with a Medtronic Model 8081 unipolar ventricular pacemaker. The patient had come to the clinic for a syncopal episode. During this recording the patient was doing an isometric exercise with his left arm.

The monitor, in this case a Medtronic programmer, adds markers to the EKG where it detects pacer spikes. These markers look like pacemaker spikes. Spike detection is controlled by the artifact setting in the programmer, and in some cases, such as in this case, over-detect ( as in the second spike seen above). The pacemaker was set to 60/min. and clearly there is a long period of imhibition of the pacemaker by the pectoral muscle with isometric exercise. This may have been related to the patient's symptoms. The problem was corrected by decreasing the sensitivity of the pacemaker.

Category

Case reports by Michael Rosengarten

Source

EKG World Encyclopedia http://cme.med.mcgill.ca/php/index.php , courtesy of Michael Rosengarten BEng, MD.McGill

Date

2012

Author

Michael Rosengarten BEng, MD.McGill

Permission

Creative Commons Attribution Noncommercial Share-Alike License

File history

Click on a date/time to view the file as it appeared at that time.

Date/TimeThumbnailDimensionsUserComment
current10:43, 21 February 2012Thumbnail for version as of 10:43, 21 February 20123,004 × 599 (646 KB)DarrelC (talk | contribs)

The following page uses this file:

Metadata