Line 1: | Line 1: | ||
== Summary == | |||
{{Information | |||
|Description = A magnet was placed over the pacemaker that was set at 30/min. and hence it is paceing at 30/min and is not affected by the native sinus beats. This and interesting case of group beating. | |||
The sinus beats are locked into the step with the ventricular pacemaker and suggests an effect of the ventricular paced complex on the sinus node, perhaps as seen with the ventriculophasic effect. | |||
|Category = [[Case_reports_from_Michael_Rosengarten|Case reports by Michael Rosengarten]] | |||
|Source = EKG World Encyclopedia http://cme.med.mcgill.ca/php/index.php , courtesy of Michael Rosengarten BEng, MD.McGill | |||
|Date = 2012 | |||
|Author = Michael Rosengarten BEng, MD.McGill | |||
|Permission = {{by-nc-sa-3.0}} | |||
|other_versions = None | |||
}} |
Latest revision as of 06:57, 21 February 2012
Summary
Description |
A magnet was placed over the pacemaker that was set at 30/min. and hence it is paceing at 30/min and is not affected by the native sinus beats. This and interesting case of group beating. The sinus beats are locked into the step with the ventricular pacemaker and suggests an effect of the ventricular paced complex on the sinus node, perhaps as seen with the ventriculophasic effect. |
---|---|
Category | |
Source |
EKG World Encyclopedia http://cme.med.mcgill.ca/php/index.php , courtesy of Michael Rosengarten BEng, MD.McGill |
Date |
2012 |
Author |
Michael Rosengarten BEng, MD.McGill |
Permission |
Creative Commons Attribution Noncommercial Share-Alike License |
File history
Click on a date/time to view the file as it appeared at that time.
Date/Time | Thumbnail | Dimensions | User | Comment | |
---|---|---|---|---|---|
current | 06:56, 21 February 2012 | 3,004 × 550 (630 KB) | DarrelC (talk | contribs) |
You cannot overwrite this file.
File usage
The following page uses this file: