McGill Case 157: Difference between revisions

From ECGpedia
Jump to navigation Jump to search
(Created page with "{{McGillcase| |previouspage= McGill Case 156 |previousname= McGill Case 156 |nextpage= McGill Case 158 |nextname= McGill Case 158 }} [[File:E0003157.jpg|thumb|600px|left|This...")
 
No edit summary
 
Line 6: Line 6:
}}
}}


[[File:E0003157.jpg|thumb|600px|left|This is a recording a patient in the pacemaker clinic. The pacer is a dual chamber and the ventricular spike is set slightly below the pacing threshold.]]
[[File:E0003157.jpg|thumb|600px|left|This is an example of time dependent capture. The native QRS complexes are small (and grow larger due to the auto gain control on this recorder), and the rhythm is probably a nodal escape with sinus arrest. The pacer is a dual chamber where the atrial lead captures but the ventricular lead captures only if it falls close to a T wave of one of the native escape beats.]]

Latest revision as of 20:21, 17 February 2012

McGill logo.png
This case report is kindly provided by Michael Rosengarten from McGill and is part of the McGill Cases. These cases come from the McGill EKG World Encyclopedia.


Previous Case: McGill Case 156 | Next Case: McGill Case 158

This is an example of time dependent capture. The native QRS complexes are small (and grow larger due to the auto gain control on this recorder), and the rhythm is probably a nodal escape with sinus arrest. The pacer is a dual chamber where the atrial lead captures but the ventricular lead captures only if it falls close to a T wave of one of the native escape beats.