McGill Case 348: Difference between revisions

Jump to navigation Jump to search
no edit summary
No edit summary
No edit summary
 
Line 6: Line 6:
}}
}}


[[File:E3481.jpg|thumb|600px|left|These are two strips from a 90 year old patient with a pacemaker.  
[[File:E0007871.jpg|thumb|600px|left|These are two strips from a 90 year old patient with a pacemaker.  
The pacemaker was past end of life and one paced beat can be seen in the top strip. Note the pacemaker was not sensing at this time. The patient's rhythm is complete heart block (note the slight variation in the pr interval which supports this interpretation)
The pacemaker was past end of life and one paced beat can be seen in the top strip. Note the pacemaker was not sensing at this time. The patient's rhythm is complete heart block (note the slight variation in the pr interval which supports this interpretation)


Line 13: Line 13:
Replacement of the pacemaker terminated the arrhythmia.]]
Replacement of the pacemaker terminated the arrhythmia.]]