McGill Case 96: Difference between revisions

From ECGpedia
Jump to navigation Jump to search
No edit summary
No edit summary
Line 6: Line 6:
}}
}}


[[File:E000795.jpg|thumb|600px|left|This cardiogram shows normal VVI pacemaker function. The patient's native EKG seems relatively normal. Of interest notice the "capture beats" where the narrow complexes that are conducted through the A/V node appear earlier than the paced beats but are narrow. This suggests that there is no retrograde conduction from the ventricle to the atrium as the sinus node seems unaffected by the ventricular pacing. This is some what similar to the case of patients with VT where capture beats are seen (very rarely)]]
[[File:E000797.jpg|thumb|600px|left|This cardiogram shows normal VVI pacemaker function. The patient's native EKG seems relatively normal. Of interest notice the "capture beats" where the narrow complexes that are conducted through the A/V node appear earlier than the paced beats but are narrow. This suggests that there is no retrograde conduction from the ventricle to the atrium as the sinus node seems unaffected by the ventricular pacing. This is some what similar to the case of patients with VT where capture beats are seen (very rarely)]]

Revision as of 01:34, 15 February 2012

McGill logo.png
This case report is kindly provided by Michael Rosengarten from McGill and is part of the McGill Cases. These cases come from the McGill EKG World Encyclopedia.


Previous Case: McGill Case 95 | Next Case: McGill Case 97

This cardiogram shows normal VVI pacemaker function. The patient's native EKG seems relatively normal. Of interest notice the "capture beats" where the narrow complexes that are conducted through the A/V node appear earlier than the paced beats but are narrow. This suggests that there is no retrograde conduction from the ventricle to the atrium as the sinus node seems unaffected by the ventricular pacing. This is some what similar to the case of patients with VT where capture beats are seen (very rarely)