File:E224.jpg: Difference between revisions

no edit summary
No edit summary
 
No edit summary
 
Line 1: Line 1:
 
== Summary ==
{{Information
|Description = This is a complex recording. The patient has a Vista DDD pacemaker which is pacing and sensing the atrium and the ventricle. Note that the paced P waves are seen clearly only in the second lead. There is artifact from the programing pulses that change the pacing rate to 60/min. with 6% rate smoothing and an upper rate of 125/mn. Note the effect of the rate smoothing which slows the pacemaker response (seen after the PVC, and later on where the PR interval is stretched)
|Category = [[Case_reports_from_Michael_Rosengarten|Case reports by Michael Rosengarten]]
|Source = EKG World Encyclopedia http://cme.med.mcgill.ca/php/index.php , courtesy of Michael Rosengarten BEng, MD.McGill
|Date = 2012
|Author = Michael Rosengarten BEng, MD.McGill
|Permission = {{by-nc-sa-3.0}}
|other_versions = None
}}