File:E0003150.jpg

Revision as of 23:43, 18 February 2012 by DarrelC (talk | contribs)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)

Original file(3,004 × 1,399 pixels, file size: 1.52 MB, MIME type: image/jpeg)

Summary

Description

In this recording we see two permanent pacemakers working at the same time. The ventricular pacemaker at a rate of 40/min., while the dual chamber pacer is set at a rate of about 80/min with an AV interval of about 160 ms. The atrial lead of the dual chamber pacemaker is capturing and sensing the atrium. The ventricular lead is not capturing the ventricle. There is no native rhythm from the patient seen here.

Category

Case reports by Michael Rosengarten

Source

EKG World Encyclopedia http://cme.med.mcgill.ca/php/index.php , courtesy of Michael Rosengarten BEng, MD.McGill

Date

2012

Author

Michael Rosengarten BEng, MD.McGill

Permission

Creative Commons Attribution Noncommercial Share-Alike License