File:E00031141.jpg

From ECGpedia
Jump to navigation Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.

Original file(3,220 × 1,815 pixels, file size: 5.55 MB, MIME type: image/jpeg)

Summary

Description

This is a cardiogram from a patient who was implanted with an Intermedics model 281-05 ventricular pacemaker 5 years before sinus node dysfunction. At the time this EKG was taken the patient was suffering from periods of atrial fibrillation and had been placed on Sotalol. This resulted in shortness of breath. The electrocardiogram is remarkable as it shows 1:1 retrograde conduction from the ventricle to the atrium, which is best seen in the lead II rhythm strip. This retrograde conduction suggested that the patient was suffering from pacemaker syndrome.

The patient was implanted with an atrial J lead to which her pacer was attached. There was resolution of her shortness of breath and so far no episodes of atrial fibrillation.

Category

Case reports by Michael Rosengarten

Source

EKG World Encyclopedia http://cme.med.mcgill.ca/php/index.php , courtesy of Michael Rosengarten BEng, MD.McGill

Date

2012

Author

Michael Rosengarten BEng, MD.McGill

Permission

Creative Commons Attribution Noncommercial Share-Alike License