File:E000795.jpg

Original file(3,004 × 1,599 pixels, file size: 4.48 MB, MIME type: image/jpeg)

Summary

Description

This cardiogram shows normal VVI pacemaker function. The patient's native EKG seems relatively normal. Of interest notice the "capture beats" where the narrow complexes that are conducted through the A/V node appear earlier than the paced beats but are narrow. This suggests that there is no retrograde conduction from the ventricle to the atrium as the sinus node seems unaffected by the ventricular pacing. This is some what similar to the case of patients with VT where capture beats are seen (very rarely)

Category

Case reports by Michael Rosengarten

Source

EKG World Encyclopedia http://cme.med.mcgill.ca/php/index.php , courtesy of Michael Rosengarten BEng, MD.McGill

Date

2012

Author

Michael Rosengarten BEng, MD.McGill

Permission

Creative Commons Attribution Noncommercial Share-Alike License