File:E0003145.jpg

Revision as of 22:47, 18 February 2012 by DarrelC (talk | contribs)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)

Original file(3,004 × 906 pixels, file size: 477 KB, MIME type: image/jpeg)

Summary

Description

This is a recording from the pacemaker clinic. The patient had a DDD pacemaker that was set to an AAI mode to check the atrial pacing threshold. This could be done as the patient had a stable ventricular escape rhythm. In this case slowing the ventricular rhythm made the P waves easier to see and the testing easier.

Category

Case reports by Michael Rosengarten

Source

EKG World Encyclopedia http://cme.med.mcgill.ca/php/index.php , courtesy of Michael Rosengarten BEng, MD.McGill

Date

2012

Author

Michael Rosengarten BEng, MD.McGill

Permission

Creative Commons Attribution Noncommercial Share-Alike License