McGill Case 108

Revision as of 08:45, 16 February 2012 by DarrelC (talk | contribs) (Created page with "{{McGillcase| |previouspage= McGill Case 107 |previousname= McGill Case 107 |nextpage= McGill Case 109 |nextname= McGill Case 109 }} [[File:E0003108.jpg|thumb|600px|left|The ...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
This case report is kindly provided by Michael Rosengarten from McGill and is part of the McGill Cases. These cases come from the McGill EKG World Encyclopedia.


Previous Case: McGill Case 107 | Next Case: McGill Case 109

The reason why the pacer was pacing at 75/min in the DDD mode was because there is a supraventricular tachycardia (note atrial activity, marked with the turquise arrows, moving through the ventricular complexes) at about 150/min and the DDD pacer was producing 2:1 A/V block. In tracing above with the pacemaker set to a VVI mode, the pacemaker no longer senses the atrium and paces at the set rate of 70/min.